I. Preparation before calibration
Equipment inspection: Confirm that the calibration bench, pressure gauge, and other tools are in good working order.
Visual inspection: Check the nameplate and valve body for rust, and ensure that all parts are complete.
II. Set pressure test
Installation: Secure the safety valve to the calibration bench.
Pressure increase: Slowly increase the pressure to 90% of the set pressure, then increase the pressure at a rate of ≤0.01 MPa/s.
Record: Record the pressure value when you hear continuous discharge of the medium. Repeat 3 times and take the average.
III. Reseating pressure test
Pressure reduction: Slowly reduce the pressure from the set pressure.
Record: Record the pressure value when the safety valve is closed.
IV. Sealing test
Pressure: Check the sealing surface below 90% of the set pressure.
Standard: When the set pressure is ≤0.3 MPa, the sealing pressure is 0.03 MPa lower than the set pressure; when >0.3 MPa, the sealing pressure is 90% of the set pressure.
V. Adjustment and Re-inspection
Adjustment: Correct the pressure value by adjusting the spring tension or the position of the counterweight.
Re-inspection: Repeat the test until it passes.
VI. Seal and Records
Seal: Seal the vessel after it passes to prevent adjustment.
Records: Archive data such as starting pressure and reseating pressure.
VII. Calibration Cycle
New Shipment: Calibrate before use if necessary.
Periodic: Perform according to the "Periodic Inspection Rules for Pressure Vessels".
VIII. Precautions
Media: Air or nitrogen is commonly used.
Safety: The vent should face away from personnel; keep your head away from the top of the safety valve.
Environment: Prevent corrosion and jamming; keep the vent clean.
